Diagnostics manufacturers who want to sell blood-bank screening assays in Brazil tend to ask me the same three things: will ANVISA accept the performance data in our instructions for use, does the INCQS pre-registration test apply to us, and can that test be run at our own factory instead of in Brazil? The questions get sharper when the assay has no approval anywhere yet, so there is no FDA or CE file to point to.
I’m Julio Martinez-Clark, CEO of bioaccess®. We help foreign device and IVD manufacturers register in Latin America and act as their local holder. Below is how I read the Brazilian rules for donor-screening immunoassays, such as Chagas and HTLV-I/II tests, under the framework ANVISA applies today.
The short answer
- Donor-screening assays are Class IV. Rule 1 in Annex I of RDC 830/2023 puts any IVD that detects a transmissible agent in blood, to decide whether it is fit for transfusion, transplant or cell administration, in Class IV. ANVISA’s own RDC 830 webinar uses HIV, hepatitis B/C, Chagas, HTLV and syphilis as its examples.
- Class IV means full registration. Under Articles 6 and 7, Classes I and II are notified and Classes III and IV are registered. A registration is valid for 10 years (Article 12).
- INCQS prior analysis applies to this category. ANVISA’s IVD classification page lists blood-bank screening tests for Chagas and HTLV, among others, as products that must go to the Instituto Nacional de Controle de Qualidade em Saúde (INCQS) for análise prévia. Approval depends on a satisfactory report.
- The analysis is done by the Brazilian public lab network, not at the factory. Article 11 §8 of RDC 830/2023 requires the report to come from a unit of the Rede Nacional de Laboratórios de Saúde Pública, under Article 16(IV) of Law 6.360/1976. ANVISA says the applicant will be asked to send samples to INCQS. I have not seen a route that accepts testing at the manufacturer’s site in its place, so plan on shipping kits.
- In-house IFU numbers are rarely enough on their own. The Class IV dossier asks for a clinical evidence evaluation report with a protocol, acceptance criteria, results and a conclusion. Performance figures from internal evaluations without a formal protocol do not usually meet that standard.
Why “in-house data” becomes the critical path
RDC 830/2023 sets out the technical dossier in Article 58, and Annex II follows the IMDRF IVD Market Authorization Table of Contents. For a screening assay, two chapters do most of the work.
Analytical performance (item VI). This covers how clinical samples were characterized and stored, accuracy, repeatability and reproducibility, analytical sensitivity or limit of detection, analytical specificity, high-dose hook effect, measuring range, cut-off definition and assay validation. Manufacturers usually have most of this already. What tends to be missing is the documentation that ties each claim to a protocol and acceptance criteria written in advance.
Clinical performance (item VIII). This needs a summary of clinical sensitivity and specificity, expected or reference values, and the clinical evidence evaluation report. For a donor-screening assay, specificity in a low-prevalence donor population matters as much as sensitivity in confirmed positives. A false positive means a deferred donor and a discarded unit.
Article 11 §4 adds a risk worth knowing about. If a filing is missing required documents, or lacks clinical evidence for a product with innovative technology or indication, ANVISA can deny it summarily rather than issue a technical query. A first-ever registration of an assay with no approval anywhere is exactly the file where I would not test that line.
My practical advice for an assay with no prior approval: run a formally protocolled clinical performance study before filing. Include well-characterized positive panels and a realistic donor population, and make sure the panels reflect the strains and subtypes Brazilian blood banks actually see. The dossier also asks for a global commercialization history. “None yet” is an acceptable answer, but it shifts more of the weight onto your own evidence.
How the INCQS step fits the timeline
INCQS is not something you book in advance like a private lab. ANVISA’s guidance says the applicant learns of the requirement through a technical query during review, and then sends samples. In practice this means:
- Keep enough kits from validated lots, with plenty of remaining shelf life, ready to ship once the query arrives.
- Budget for import logistics into Brazil under your holder’s company authorization. Kits that expire in customs are a common and avoidable loss.
- Treat INCQS as sequential to the dossier review, not parallel to it. The registration is granted only after a satisfactory report.
The other gates for a foreign manufacturer
- A Brazilian holder. Article 14 requires a declaration from the legal manufacturer, consularized or apostilled, authorizing the applicant company to represent and sell the product in Brazil.
- ANVISA GMP certification. Class III and IV registrations need an ANVISA Certificate of Good Manufacturing Practice (RDC 665/2022), or proof that the certification request has been filed. The filing receipt lets review start, but ANVISA cannot grant the registration until the certificate is published (Article 14 §§1-2). For a factory ANVISA has never inspected, this is often the longest lead item, so start it first.
- Stability. Shelf life has to come from real-time data, or from accelerated data with the real-time study under way, on at least three lots (item VII). For screening reagents shipped long distances, plan transport stability studies too.
- Labeling and IFU in Portuguese, following Articles 46 to 50.
What I would do first
- Confirm the Class IV classification and the technical name in ANVISA’s IVD list, so the file is scoped correctly from day one.
- Choose the Brazilian holder and file for the ANVISA GMP certificate, because that clock is the longest.
- Do a gap review of the analytical and clinical evidence against Annex II. If the clinical evidence report cannot be written from existing data, design the clinical performance study now. Running it in Brazil, with local donor and confirmed-positive panels, makes it easier to defend to reviewers.
- Reserve validated lots for INCQS and document their stability.
Frequently asked questions
What risk class are Chagas and HTLV blood donor screening tests in Brazil?
Class IV under Rule 1 of Annex I of ANVISA RDC 830/2023, because they detect transmissible agents to decide whether blood is fit for transfusion. Class IV IVDs need full registration.
Is INCQS testing mandatory for blood-bank screening IVDs?
Yes. ANVISA lists blood-bank screening tests, including Chagas, HTLV, HIV, hepatitis B and C, syphilis and the ABO, Rh and irregular-antibody immunohematology reagents, among the IVDs that must go to INCQS for prior analysis. Dengue, chikungunya and SARS-CoV-2 self-tests are also on the list.
Can the INCQS analysis be done at the manufacturer’s facility abroad?
RDC 830/2023 requires the report to come from a unit of Brazil’s national public health laboratory network, and ANVISA instructs applicants to send samples to INCQS. Plan on shipping kits to Brazil.
Will ANVISA accept performance data from the manufacturer’s internal evaluations?
Only if it supports a clinical evidence evaluation report with a protocol, acceptance criteria, results and a conclusion. Internal evaluations without a formal protocol usually need to be supplemented with a protocolled clinical performance study.
Do investigational IVDs used in a Brazilian clinical performance study need registration?
No. Article 2 §1 of RDC 830/2023 exempts IVDs intended for clinical investigation from notification or registration, provided the legal requirements for the research are met. They cannot be sold or used for any other purpose.
For more on Brazil, see our ANVISA registration roadmap and the Brazil import permit guide. If you are scoping a Brazilian launch for a diagnostics line, our market access team can map the classification, holder model, GMP certificate, INCQS and evidence gaps before you commit budget.
This article reflects ANVISA rules as published at the time of writing (November 2026): RDC 830/2023 (in force since 1 June 2024) and ANVISA’s IVD classification guidance. It is general information, not regulatory advice for a specific product.